4. Evaluation
The measured value constitutes
the resistance R [M] of the liquid
between the electrodes (in the
annular gap). The cell constant
which depends on the design and
dimensions of the measuring cell
is 7.55 10
-3
cm
-1
.
The specific resistance, , and the
specific conductivity are
calculated from the following
equation:
= R • 132,5 cm
or [M cm] =
R [M]
7,55 • 10
-3
cm
-1
[S • cm
-1
] = 7,55 • 10
-3
cm
-1
R [M]
with 1 = 1S
-1
= 1
1S
If the measurement result is
required for internal quality control
only, e.g. for entries in a check-
card, conversion is not necessary.