Configuring BYK LC 3
10
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
! Switch off the measuring instrument.
! While pressing the <M> key, press <>.
! Select the required Baud rate with <> <>.
Selection: 1200, 2400, 4800, 9600 Baud.
! Then press the <RUN/ENTER> key.
! Set the required calibration interval with <> <>.
Selection: 1... 180... 999 d.
! Then press the <RUN/ENTER> key.
! Switch the automatic measuring range selection AutoRange on/off with
<> <>.
Selection: YES (On), no (Off).
! Then press the <RUN/ENTER> key.
! Switchover the reference temperature of the conductivity with <> <>.
Selection: 25 °C (Tref25) and 20 °C (Tref20).
! Then press the <RUN/ENTER> key.
! Switch the units of the conductivity display on/off with <> <>.
Selection: S/cm, M
cm.
! Then press the <RUN/ENTER> key.
! Select the date and time step-by-step with <> <>. In between, press
the <RUN/ENTER> key each time.
! When the last <RUN/ENTER> key has been pressed, the instrument
changes to the last active measuring mode.
Baud
0
40
8
Time
0
d
T
3
n
1
8
1
S
E
Y
ARng
n
r
A
Tref25
2
5
t
d
n
o
C
cm
S/
Day.Month
4
00
9