22
File Display Option
Measure Gloss Std
Press Operate Button
Std.No.
:
976802
Calibrate
Use the
buttons to move the cursor to the “Option” menu.
From this menu, use the
button to place the cursor on the “Calibr.” option.
Press the Operate button. When the submenu
opens, use the
button to move the cursor to the “Calibrate” option.
Now press the Operate button, and calibration will
begin. The instrument will first call for the black
standard. Place the instrument on the black
standard and press the Operate button.
A beep will indicate that the measurement has been
completed.
Now place the instrument on the white standard and
press the Operate button.
For gloss calibration, place the instrument on the
high-gloss standard and press the Operate button.
If calibration was performed correctly, the message
Calibration OK
will appear on the display. The instrument returns to
the previous measurement mode.
Use the test with the green standard again to
determine whether the calibration was performed
correctly.
std
std
std
File Display
Color sys
Index
Illum/Obs
chk green
Setup
Configure
f
f
f
f
f
Option
Calibr.
Calibrate
f
Gloss
f
File Display
Option
Measure White
Press Operate Button
Std No 638444
File Display
Color sys
Index
Illum/Obs
Setup
Configure
f
f
f
f
f
Option
Calibr.
f
Gloss
f
File Display Option
Measure Black Trap
Press Operate Button