44
The battery or the Change or recharge
rechargeable bat- the battery (chap. 4.1).
tery is discharged.
The capacity of the If applicable, transfer
memory in STATIS- the memory contents
TIC MODE is ex- to a PC and clear the
hausted (999 loca- memory (chapter 9.3).
tions).
The calibration was Put unit into the correct
started on the wrong holder and, if neces-
standard. sary, restart calibration
(chapter 7).
Inadequate light in- Repeat calibration. If a
tensity during cali- fault is detected, please
bration (lamp aging contact our Customer
or operator error). Service.
One of the elec- Please contact our
tronic components is customer service.
defective.
Lower limit of Hlog > 10 units min.
measuring range (chapter 10).
exceeded.
Lower limit of Gloss > 10 units min.
measuring range (chapter 10).
exceeded.
15 Error Messages
>>LOW BATTERY!<<>>LOW BATTERY!<<
>>LOW BATTERY!<<>>LOW BATTERY!<<
>>LOW BATTERY!<<
MEMORY OVERFLOWMEMORY OVERFLOW
MEMORY OVERFLOWMEMORY OVERFLOW
MEMORY OVERFLOW
--
--
-
>>
>>
>
modemode
modemode
mode
WRONG HOLDERWRONG HOLDER
WRONG HOLDERWRONG HOLDER
WRONG HOLDER
CONTROL LAMPCONTROL LAMP
CONTROL LAMPCONTROL LAMP
CONTROL LAMP
UNITUNIT
UNITUNIT
UNIT
NEEDS SERVICENEEDS SERVICE
NEEDS SERVICENEEDS SERVICE
NEEDS SERVICE
< 10< 10
< 10< 10
< 10
RANGERANGE
RANGERANGE
RANGE