BYKLC4 What to do if...
47
6 What to do if...
Error message
0FL
0FL0FL
0FL, UFL
UFLUFL
UFL
Error message
Err3
Err3Err3
Err3
Error message
Err4
Err4Err4
Err4
Error message
Err5
Err5Err5
Err5
Cause Remedy
The measured value lies outside
the measuring range
Sensor module defective Replace sensor module
Cause Remedy
Erroneous calibration
The data of the sensor module is
outside the specified characteris-
tics.
Recalibrate
Reset to default setting (see
section 4.12)
Sensor module contaminated Clean and replace sensor
module if necessary
Moisture in the plug connection Dry plug and socket
Calibration solution not suitable Check calibration solution
Cause Remedy
No sensor module Connect sensor module
Check the plug connection
Switch the meter off and then on
again.
Perform a processor reset (see
section 6.1)
Replace sensor module
Cause Remedy
Sensor module defective
or
Wrong sensor module
Check the plug connection
Switch the meter off and then on
again.
Perform a processor reset (see
section 6.1)
Replace sensor module