BYKLC4 Operation
27
4.5.1 Selecting the temperature compensation
Setting menu for
the temperature
compensation
1 Open the setting menu for the cell constant with <C>.
The display shows the setting last selected.
2 Open the setting menu for the temperature compensation with <T>.
The display shows the setting last selected.
3 If necessary, press <ß_> repeatedly until the nLF, Lin or OFF display
appears.
– nLF = nonlinear temperature compensation
– Lin = linear temperature compensation
– OFF = temperature compensation switched off
4 – Confirm the setting with <g>.
in the nLF menu --> back to the measuring mode
–in the Lin menu --> the menu for setting the temperature coef-
ficient opens up (see section 4.5.2)
–in the OFF menu --> back to the measuring mode
The displayed setting is used immediately.
CAL
ESC
S/cm
K
2
Cell
1
cm
Sal
mg/l
Tref20Tref20
nLF
TDS
MEM
CAL
C
m
RCL
S
SAL
O
2
TC
RESET
Cell
LAP
mbar
STO
LoBat
Day.Month
h:min min:s
Ident
Lap
CLR
M
cm
k
cm
r 2t
L i n
5
Tref25
TC
Lin
<ß_>
ESC
CAL
S/cm
02rt
Tref25
Cell
Lin
1
cm
0
Sal
mg/l
Tref20Tref20
nLF
TDS
MEM
CAL
C
m
RCL
S
SAL
O
2
TC
RESET
Cell
LAP
mbar
STO
LoBat
Day.Month
h:min min:s
Ident
Lap
CLR
M
cm
k
cm
K
O F F
TC
<ß_>
<ß_>
S/cm
CAL
ESC
Lin
Cell
1
cm
0
Sal
mg/l
Tref20Tref20
TDS
MEM
CAL
C
m
RCL
S
SAL
O
2
TC
RESET
Cell
LAP
mbar
STO
LoBat
Day.Month
h:min min:s
Ident
Lap
CLR
M
cm
k
cm
K
r 2t
n L F
5
Tref25
TC
nLF
Comentarios a estos manuales