Normalization performed by mistake on calibration
standard instead of on uncoated specimen.
Perform normalization on uncoated specimen.
Er14 Internal instrument error: Not able to compute
master characteristic. The original master
characteristic is retained.
Repeat master calibration.
If this error occurs repeatedly:
Contact customer service.
Er15 Not able to save the master characteristic.
Probe may be defective.
Contact customer service.
Er17 Not enough readings. Thus, not able to finish a
block (applies to measurements w. fixed block
limits).
Make additional measurements until the first
block can be finished. Then generate a new final
result.
Er22 Internal instrument error. Error accessing the
measurement memory or. EEPROM.
Contact customer service.