13
5.3.1 Triggering the mean value transmission manually
Prerequisite: Settings of service menu must be “bL = 1” e.g. mean value transmission is active.
See chapter 13 “Service menu, menu #1”, menu point “Block length”.
1. Make desired number of measurements.
2. Set the display menu with [] or [] to show "RES”. Press button [OK]. The mean value of the readings in the
memory are then radioed to the PC. This is indicated by the flashing radio tower symbol.
5.3.2 Sending the mean value automatically
Prerequisite: Service menu settings of “bL” must be between 2 and 20.
The set value indicates the number of readings, which are collected in one block. See chapter 13 “Service menu”.
1. Make desired number of measurements.
2. Indepentently of the pre-set number of readings/block, a block mean value is radioed and a new block will be
begun.