3.5 Place the Pencil Tester Body on end, onto a
rm horizontal surface (arrow pointing
downwards.
3.6 Slowly slide (care should be taken not to chip
the pencil edges) the prepared pencil into the
hole adjacent to the arrow until it rests on the
same rm horizontal surface as the pencil
tester body. Turn the pencil clamping knob
clockwise to lock the pencil in position.
Carefully lift the pencil tester off the surface,
ensure pencil is set rm and is free from
damage to edges. Check the bubble level
to conrm the pencil has been properly
positioned.
always ensure the position of the pencil
is the same, thereby applying the consistent mass
to the coating under test.
Place the pencil tip gently onto the scales and read
the applied mass, value to be 750 (+10/-10) grams.
If the mass is outside of the specication, move the
slide weight forward to increase the mass or back to
decrease the mass.
3.7 Carefully place the prepared pencil tester,
wheels rst onto the surface to be tested,
gently bring the pencil tip into contact with
the coated surface to be tested, ensure no
chipping occurs.
3.8 Immediately after the tip of the pencil rests
on the coated surface, place thumb and
forenger on the pencil tester, centre portion
of each wheel, push the instrument in the
direction of the arrow for a suggested distance
of 7mm.
3.9 Inspect the coating after a specied period of
time with the naked eye check for marking of
the surface.
3.10 If no marking has occurred, repeat the test
(3.1 to 3.8) on a new portion of undisturbed
Comentarios a estos manuales