Altana Micro-TRI-Gloss μ Manual de usuario Pagina 59

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 174
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 58
58
English
Memory full Transfer the content of memory to a PC and then
delete the contents of memory.
Reference memory full A maximum of 50 references can be saved. It
may be necessary to delete old references.
You will also find an error number for the following messages in the
Calibration/Status menu to provide support for diagnostics:
Tolerance Generally occurs only with major changes in
Error 01 climatic or weather conditions. The deviation was
successfullycompensated for by calibration and
correct measurements are still possible. You
should still recalibrate the device as soon as it is
operating in normal climatic conditions again.
However if a change in climate cannot be
considered as the cause of the problem, you
should check whether the standard is clean.
Please call Customer Autodiagnosis has determined an impermissible
Service (invalid) deviation in the measurement
signal that cannot be remedied by recalibrating.
Error 02 Generally occurs when there is a significant
amount of dirt or dust on the standard or optics.
First try to clean the standard. You should only
have the optics cleaned by our Customer Service
department, for example as part of a yearly
recertification.
Error 03 Defect in the electronics or operating error. First
check whether the standard is clean and whether
the device is properly snapped into the holder.
Error 04 Defect in the lamp or electronics.
Error 05 Defect in the electronics.
Infi Wrong basis metal, measurement range
exceeded or calibration error; recalibrate, if
necessary.
22 Errors and warning messages
Vista de pagina 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 173 174

Comentarios a estos manuales

Sin comentarios