Altana Micro-Gloss Manual Manual de usuario

Busca en linea o descarga Manual de usuario para Instrumentos de medición Altana Micro-Gloss Manual. ALTANA Micro-Gloss Manual User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - Measure what you see

Additives & InstrumentsA member ofMeasure what you see.micro-glossManual

Pagina 2

9Power supply2. Power supplyBefore operating the instrument for the first time,please read the operating instructions and payattention to the safety

Pagina 3

10Power supply2.2 Changing the batteryTo insert or change the battery open the batterycompartment. The easiest way to do this is byturning the cover w

Pagina 4 - Table of content

11ControlsMeasurement unit and protective holder1 Mode scroll wheel: used to turn the unit on and for menu selection2 Display for user guidance and di

Pagina 5

12ControlsThe basic system consists of the measuring deviceand the protective holder.The protective holder is used for calibration and tostore the mea

Pagina 6

13Getting started4. Getting started4.1 Turning on the unit and measuringTo turn on the unit, press the mode scroll wheel.Information on the date and l

Pagina 7

14Getting started4.2 NavigationAll control functions are controlled by the modescroll wheel. Pressing the wheel causes a menu toappear in the display.

Pagina 8

15Getting started4.3 Change names/numbersFor some functions, you can enter or change thedate or name. The arrow pointing upward marks theposition that

Pagina 9

16Getting startedMeasurement without statistical evaluation.Multiple measurement with statistics.Continuous measuring with adjustable interval.Measuri

Pagina 10 - 2. Power supply

17Calibrate5. CalibrateThe holder with the integrated glass standard isused for calibration. Always keep the measurementunit in the holder. This prote

Pagina 11

18Calibrate5.2 CalibrateYou should recalibrate the device if ambientconditions have changed. This applies especiallywhen changing location if major ch

Pagina 12 - 3. Controls

1micro-glossManualPatent pending 260 020 398 E 1008BYK-Gardner GmbHLausitzer Str. 8D-82538 GeretsriedGermanyTel. 0-800-gardner(0-800-4273637)+49-8171

Pagina 13

19After the zero setting has been performed, thedisplay AIR will appear. To proceed with this, holdthe instrument in midair and press the operatebutto

Pagina 14 - 4. Getting started

20CalibrateA warning message appears. You can cancel thisprocess by pressing the operate button.If you press the scroll wheel, you will continue witht

Pagina 15

21CalibrateYou can set the scales of the measurement valuesfor gloss and thickness separately.5.2.5 Scale GlossYou can use the Scale menu option to sw

Pagina 16

22Calibrate5.3 Calibrating standardsTo ensure exact calibration, only original standardsfrom the manufacturer should be used.These are calibrated agai

Pagina 17

23Measurement techniques6. Measurement techniquesIn accordance with the standard, the reflectometervalue is related to a black glass standard at adefi

Pagina 18 - 5. Calibrate

24Measurement techniques6.2 Anodized aluminum and other metal surfacesThe measuring unit is equipped with an extendedmeasuring range for measuring sam

Pagina 19

256.3 Measurement of film thicknessThe film thickness sensor functions according to themagnetic method (Fe) or the eddy current method(NFe). Therefore

Pagina 20

267. Measurement ModesYou can select different types of measurement inthe Mode menu. The mode that is activated isidentified by a check mark.7.1 Sampl

Pagina 21

277.2 StatisticsYou can make multiple measurements with eachsample in Statistics mode. These measurementswill be evaluated statistically and displayed

Pagina 22

28Measurement ModesNote:(micro-TRI-gloss µ only)When STATISTICS is activated, all film thicknessvalues are stored in the Memory independently ofthe so

Pagina 23

2Dear customer,thank you for having decided for a BYK-Gardnerproduct. BYK-Gardner is committed to providing youwith quality products and services. We

Pagina 24 - 6. Measurement techniques

29Measurement Modes7.2.2DisplayIn the Statistics measurement display, you canassign the following data freely to three columns:Value:Last value to be

Pagina 25

30Measurement Modes7.2.3Exit blockThis function terminates the block before it reachesthe required number of measurements n. It is usefulif you have

Pagina 26

31Measurement Modes7.3 ContinuousYou can use this function to perform up to 99measurements at an adjustable measurementinterval. This is helpful when

Pagina 27 - 7. Measurement Modes

32Measurement ModesThe measuring interval can be changed before asequence is started. Therefore press the modewheel to open the Continuous submenu.The

Pagina 28

338 Geometry/SensorIn this menu, you can select the geometry for thegloss measurement, just as the sensor for the filmthickness measurement (at micro-

Pagina 29

34Geometry/SensorCombiWhen changing the substrate, the sensor will beswitched over automaticallyChoose the desired sensor with the scroll wheel.Press

Pagina 30

359. MemoryTo save measurement values, you must activate theMemory function before measuring or else select orcreate a memory. Up to 999 measurements

Pagina 31

369.3 Create memoryUsers can set up their own memory areas with thisfunction. Select the required geometry before youactivate this function. Then you

Pagina 32 - 7.3 Continuous

37MemoryThe values of the first measurement appear in thedisplay. The sample name is displayed in thehighlighted field.Turning the wheel switches the

Pagina 33 - 7.4 Basic mode

38Difference measurement and Pass/Fail10. Difference measurement and Pass/FailYou can compare the readings of samples with thevalue of a previously me

Pagina 34 - 8 Geometry/Sensor

3EnglishTable of contentTable of content1. System description and Delivery notes ... 72. Power supply ...

Pagina 35

39The measured standard values are saved as thetarget values. At the same time, Differencemeasurement is turned on and the measuredstandard is activat

Pagina 36 - 9. Memory

40Difference measurement and Pass/Fail10.4 Create standardStandards can also be saved by entering the targetand limit values with the scroll wheel. Mo

Pagina 37

41Difference measurement and Pass/Fail10.5 Change standardYou can use this function to change target valuesand limit values of saved standards. You ca

Pagina 38

4211. SetupYou can make general settings in the Setup menu,for example Language or Display time.11.1 Bluetooth®You can use this menu option to turn th

Pagina 39

4311.5 LanguageYou can use this menu to select the displaylanguage.Use the scroll wheel to move the mark to thedesired language and press the wheel.11

Pagina 40

44Interface12. InterfaceThe measurement device is equipped with aninterface that allows direct communication with aPC.Measurement data can be transfer

Pagina 41

4512.1 InstallationBluetooth® :Please refer to the separate Bluetooth® InstallationGuide.Software easy-link and USB drivers:Insert the easy-link CD i

Pagina 42

46Standards13. StandardsISO 2813 Paints and varnishes - Determination of speculargloss of non-metallic paint films at 20°, 60° and 85°ASTM D 523 Stand

Pagina 43 - 11. Setup

47ISO 2178 Non-magnetic coatings on magnetic substrates -Measurement of coating thickness - MagneticmethodISO 2360 Non-conductive coatings on non-magn

Pagina 44

48Technical data14. Technical dataGeneral technical dataTemperature range +15 °C to +40 °C (60°F to 104°F) for operation - 10 °C to +60 °C (-14°F to 1

Pagina 45 - 12. Interface

4Table of content7.2.3 Exit block ...308.2.4 Delete

Pagina 46

49Technical dataFilm thickness measurement:Substrate Fe: magnetic, e.g. ironNFe: non magnetic, e.g. aluminiumSonde One pointMeasurement range 0...500

Pagina 47 - 13. Standards

50Errors and warning messages15. Errors and warning messagesMemory full Transfer the content of memory to a PC and thendelete the contents of memory.R

Pagina 48

51Errors and warning messagesError Thickness Operating error: improper application, raisingbefore measurement is complete, or calibrationerror. Repeat

Pagina 49 - 14. Technical data

5216. Cleaning and maintenance• Do not insert any objects into the measurementaperture for cleaning. The instrument could getdamaged - affecting a pr

Pagina 50

5317. Service and CertificationServiceBesides the repair of your instrument we offer thefollowing additional services:First diagnosis on the telephon

Pagina 51

54Service and CertificationService Centers for BYK-Gardner productsGermanyBYK-Gardner GmbHLausitzer Strasse 882538 GeretsriedGermanyPhone:+49-8171-34

Pagina 52

5518. CopyrightThis instruction manual is an important part of this instrument. It contains es-sential information about setting up, placing in servic

Pagina 54

5Table of content12. Interface...4412.1 Installation ...

Pagina 55

6System description and Delivery notes1. System description and Delivery notesMeasurement units of the micro-gloss family can beused to determine the

Pagina 56 - 18. Copyright

7Instrument type:micro-gloss 20° 4440micro-gloss 60° 4442micro-gloss 85° 4444micro-gloss 45° 4454micro-gloss 75° 4456micro-TRI-gloss 4446micro-TRI-glo

Pagina 57 - 260 020 398 E 1008

8Recommended accessories:Checking standards for control of test equipmentfor micro-gloss 20° 4422for micro-gloss 60° 4462for micro-gloss 85° 4487for m