Advanced Operations
7. Remove any present cuvettes from the cuvette
compartment.
8. Close the cuvette compartment cover.
9. Press Start.
10. Follow the instructions on the screen and insert the different
filters one after another in the sequence shown.
11. Press Next after inserting a filter.
After the last reading the results are displayed.
12. Press PC & Printer icon to send the data to a USB memory
stick, PC or to a printer.
The files will be stored automatically as CSV files (Comma
Separated Value). The file name will be formatted as
"Verification.csv".
6.1.4 AQA Analytical Quality Assurance
6.1.4.1 Standard Configuration
Analyzing an AQA standard solution enables you to check your
working procedure, photometer and accessories.
1. Under "System Check", select AQA,
or
press AQA in the toolbar on the right of the screen.
A list of standard solutions opens.