Altana Haze-Gard Plus Calibration instructions for Taber Manual de usuario Pagina 2

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 2
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 1
haze-gard plus
Calibration Procedure with Taber Abrasion Holder
Calibrate the instrument without the Taber Abrasion Holder.
Install the Taber Abrasion Holder in front of the sphere entrance.
Calibrate the instrument once more, but only dark- and bright calibration.
When you are asked to calibrate Clarity, please press the ”set-up“ button.
Perform measurements without a sample in Haze, Transmission and Clarity.
The result should be: H = 0.00, T = 100 and C = 100.
If the Taber Abrasion Holder is removed, calibrate the instrument again.
Note:
When working with the Taber Abrasion Holder, only measure Haze or Transmission.
Clarity measurements are not possible with the holder due to changed
optical conditons.
BYK-Gardner USA
9104 Guilford Road
Columbia, MD 21046
Phone 301-483-65-00
Fax 301-483-65-55
http://www.byk.com/instruments
238 019 149 ED 1007
Vista de pagina 1
1 2

Comentarios a estos manuales

Sin comentarios