Altana Haze-Gard Dual Manual de usuario Pagina 52

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 164
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 51
51
English
No statistic
values stored
SET UP
Printer
not ready
Escape:
Invalid date
Cause
The current statistics memory in the
measurement device does not
contain any data.
Cause
The connected printer is not re-
sponding. This error only occurs
when statistics data is being prin-
ted. The printer is not polled
during online mode.
Remedy
Check that the printer cable is
correctly connected!
Is the printer switched on?
Is there paper in the printer?
Is the printer expecting a
postscript file?
Cause
A date earlier than 1980 was input
during SET-UP.
Remedy
The date input was ignored and the
old date retained. Enter the correct
date.
16 Error Messages
Vista de pagina 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 163 164

Comentarios a estos manuales

Sin comentarios