3.0 Cupping Test
Carry out the cupping test as follows:
Insert a test panel (max. thickness
1 mm) into the the cylinder opening.
Turn the eccentric clamping ring
(10) to slightly clamp the test panel.
Press the START button
K and
closely watch the sample through
the stereo microscope.
With the fi rst visible crack of the
paint lm on the test panel press the
STOP button and read the cupping
value from the digital display.
Press the REVERSE button
L to
move back the indenter.
After the indenter reaches the initial
position, loosen the eccentric
fastening device and take out the
test panel.