
17
Calibrate
7.2 Calibration Notes
• The standards should be cleaned periodically.
For cleaning procedure see section "Cleaning
and Maintenance“. Please make sure that the
standards are not scratched.
• Do not move the instrument while taking a
calibration measurement. If motion is detected,
an error message will be displayed and
calibration is aborted.
• When moving from cold to warm environment,
there is a danger of condensation. For this
reason, you should wait for an appropriate
amount of time to allow the optical components
to adjust before calibrating and using the unit.
Comentarios a estos manuales