Altana Wave-Scan ROBOTIC Manual de usuario Pagina 11

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 45
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 10
10
English
Power Supply and Communication
Robotic PC oder
SUBD9pin RS422
Male Converter
Pin Signal Signal
1 RxD+ (B) TxD+ (B)
2 RxD- (A) TxD- (A)
3 TxD+ (B) RxD+ (B)
4 TxD- (A) RxD- (A)
5 GND GND [*]
6 U_in
7 GND
8 U_in
9 GND
[*] optional;
U_in 6,8 and GND 7,9 for power
supply.
Kabel:
min. 8 wires, twisted-pair; 0,20mm²
RxD+ ; RxD- und TxD+ ; TxD- and
two pairs for U_in ; GND
e.g. LifYCY (4*2 )*0.20 or (6*2 )*0.20
(e.g. http://www.metrofunk.de/)
Power Supply:
Input: 24VDC; Max. 0,5A
4 Interface
Attention: Please heed tension
and pin assignment precisely.
The instrument can be damaged
without functional connection. No
product liability and guarantee claims
can be put forward if the instrument is
not properly power supplied.
Vista de pagina 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 44 45

Comentarios a estos manuales

Sin comentarios